Invitrogen™ DynaMag™-96 Bottom Magnet
Description
Includes
One magnet
Magnet pulls beads to the bottom of each well. Compatible with PCR-strips and 96-well plates (200 μL) non-skirted and half-skirted and replaces the Dynal MPC-96B. When used in combination with the DynaMag-96 Side on an automation platform, the two magnets allow for significant volume changes through a workflow.
- High wash volumes—DynaMag-96 Side
- Low Elution Volumes—DynaMag-96 Bottom
- Concentrates Beads at the bottom of each well
- Features 96 round magnets (one centered at the bottom of each well: to ensure identical sample handling at each position)
- Allows for concentrating the sample by eluting in tiny volumes (approx. 10μL)
Order Info
Shipping Condition: Room Temperature
Specifications
Specifications
| Description | Magnet |
| Content And Storage | Contains one magnet. Store at room temperature. |
| For Use With (Equipment) | DynaMag™ |
| Shipping Condition | Room Temperature |
| Product Line | DynaMag |
| Quantity | 1 each |
Frequently Asked Questions (FAQs)
The DynaMag-2 Magnet is 3,500-3,700 minimum Gauss level and the DynaMag-Spin is 3,000-3,500 minimum Gauss level.
1 Tesla is equivalent to 10,000 Gauss. .
Spray and/or wipe the DynaMag magnet unit with one of the following cleaning agents:
- 70% (v/v) isopropyl alcohol
- 1% (v/v) sodium hypochlorite solution (bleach)
- 0.1 N HCl solution
Do not submerge the magnet unit in aqueous solutions and avoid prolonged exposure to water or aqueous solutions. Clean with a damp cloth and mild detergent when exposed to harsh solvents. Do not autoclave the magnet unit as heating the magnet to above 70 degrees C can destabilize the magnetic material, resulting in loss of magnetic strength.
The DynaMag-2 Magnet can hold 16 standard 1.5 mL or 2 mL microcentrifuge tubes, while the DynaMag-Spin Magnet holds up to 6 standard 1.5 mL microcentrifuge tubes.
Superparamagnetic means that the Dynabeads magnetic beads exhibit magnetic properties when placed within a magnetic field, but have no residual magnetism when removed from the magnetic field.
This means that your targeted cells, proteins, or nucleic acids are only subjected to magnetic forces during the time the beads are on the magnet. The beads do not aggregate, but remain evenly dispersed in suspension.
Safety and Handling
For Research Use Only. Not for use in diagnostic procedures.