Learn More
Description
The Thermo Scientific™ Orion™ Conductivity electrode 013016MD complies with SM 2510B; D1125-95, -99 (A) for Drinking water measurements and USP<645> Measuring the Conductivity of Pure Water.
Features
These probes use two cell electrodes to measure low-, standard-, and high-conductivity samples based on the cell constant. They are a versatile option for precise measurements in the lab and field.
013016MD is ideal in low ionic strength solutions, such as deionized water and ultrapure water when paired with a flow-cell assembly . 018020MD is good for chemically reactive conductivity samples, including concentrated acids and bases. 011050MD and 011510MD epoxy/graphite cells are durable, chemical resistant, and measure a wide range of conductivities in the lab or field.
2-cell conductivity electrodes are compatible with Thermo Scientific™ Orion Star™ , Lab Star™, Pro Star™, Versa Star Pro™ bench and portable conductivity meters. Paired together the instruments can be used to generate accurate, repeatable readings for conductivity, resistivity, TDS and salinity.
Solutions required:
Standards
- 011008 100 μS/cm conductivity/TDS standard, 5 x 60 mL bottles
- 01100910 147 μS/cm conductivity standard, 10 pouches
- 011007 1413 μS/cm conductivity/TDS standard, 5 x 60 mL bottles
- 01100710 1413 μS/cm conductivity/TDS standard, 10 pouches
- 011006 12.9 mS/cm conductivity/TDS standard, 5 x 60 mL bottles
- 01100610 12.9 mS/cm conductivity/TDS standard, 10 pouches
- 011005 111 mS/cm conductivity standard, 5 x 60 mL bottles
- 01100510 111 mS/cm conductivity standard, 10 pouches
Conditioning solution
- 011001 Conditioning solution for 011050MD conductivity cells
Accessories
- 013017 Replacement flow cell
Items included with
011050MD
- 011007 1413 μS/cm conductivity/TDS standard, 60 mL 1 bottle
- 011001 Conditioning solution, 60 mL 1 bottle
011510MD
- 011006 12.9 mS/cm conductivity/TDS standard, 60 mL 1 bottle
- 011007 1413 μS/cm conductivity/TDS standard, 60 mL 1 bottle
013016MD
- 011008 100 μS/cm conductivity/TDS standard, 60 mL 1 bottle
013017 Replacement flow cell
Warranty period
- 12 months from date of purchase
Specifications
Specifications
| Cell Constant | 10 cm^-1 |
| Type | 2-Electrode Conductivity Cells |
| Connector Types | Mini-DIN |
| For Use With (Equipment) | Orion Star Series Meters |
| Description | Orion 2-Cell Conductivity Electrodes |
| Material | Glass, Platinum |
| Measurement Ranges | 10 μS/cm to 2000 mS/cm |