Learn More
Thermo Scientific™ Orion™ Versa Star Dissolved Oxygen / RDO Measurement Module
Add dissolved oxygen and temperature measurement capabilities to your Versa Star meter using the Thermo Scientific™ Orion™ Versa Star Dissolved Oxygen / RDO Measurement Module
Supplier: Thermo Scientific™ VSTARRD
Description
Add dissolved oxygen and temperature measurement capabilities to your Thermo Scientific™ Orion™ Versa Star meter using the Thermo Scientific™ Orion™ Versa Star Dissolved Oxygen / RDO Measurement Module. Each Orion™ Versa Star meter has four channels that can be customized with interchangeable measurement modules. The Orion™ Versa Star RDO/DO module offers the flexibility to use Orion™ RDO optical or polarographic probes with automatic sensor recognition and calibration using water-saturated air, air-saturated water, custom value using Winkler titration and zero point options.
- Measure dissolved oxygen in percent saturation or mg/L with temperature using Orion RDO optical DO probes or Orion polarographic DO probes and automatic sensor recognition
- Calibrate dissolved oxygen using water-saturated air, air-saturated water, custom value (Winkler titration) and zero point options
- Automatically compensate dissolved oxygen readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure readings
- Customize measurements to show calibration history or simplify to show an extra large dissolved oxygen reading
- Module offers a 3-year replacement warranty
Disclaimer to internal research use restriction: Any restrictions on the purchaser to utilize the product(s) for internal research purposes only does not apply to the product(s) on the web pages associated with this statement.
Specifications
±0.2 mg/L or ±2% of reading, whichever is greater | |
±6mmHg | |
Set point alarm, high/low limit alarm, calibration due alarm | |
Water-saturated air, air-saturated water, manual (Winkler titration), zero point | |
12 x 3 x 9.5 cm (4.7 x 1.2 x 3.8 in.) | |
Versa Star RDO/DO module and printed quick start guide | |
Dissolved oxygen as % saturation or mg/L with temperature and barometric pressure | |
Orion Versa Star RDO Optical/Dissolved Oxygen/Temperature Measurement Module | |
0.1, 1% Saturation | |
0.1mmHg | |
±2% Saturation | |
0.00 to 50.00mg/L | |
0.01, 0.1 mg/L; 0.1, 1 % saturation | |
0.1°C | |
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting | |
-5°C to 105deg;C | |
5°C to 45°C | |
0.5 lb. | |
0.2 kg | |
±2 % saturation ≤ 200 % saturation, ±10 % saturation > 200 % saturation |
±0.1mg/L up to 8mg/L, ±0.2mg/L from 8 to 20mg/L, ±10% of reading up to 50mg/L | |
±0.1°C | |
400 to 850 mmHg, automatic using built-in barometer or manual entry | |
Orion Versa Star RDO/DO/temperature measurement module | |
9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature) | |
IP-54 | |
10 per channel with password protection | |
0.0 to 500.0 % saturation | |
400 to 850mmHg | |
RDO optical dissolved oxygen probe, polarographic dissolved oxygen probe (purchase separately) | |
0.00 to 90.00 mg/L | |
5 to 85 %, non-condensing | |
0.01, 0.1 mg/L; 0.1, 1 % saturation | |
0 to 45ppt | |
41°F to 113°F | |
1 point temperature offset calibration | |
Automatic with ATC temperature probe | |
3 year meter replacement | |
0.1, 1 % saturation | |
0 to 600% Saturation |
Your input is important to us. Please complete this form to provide feedback related to the content on this product.