Learn More
YSI™ TruLab™ pH/ISE 1320 Dual Channel Laboratory ISE Benchtop Meter
Shop All Yellow Springs Instrument (YSI) ProductsDescription
Includes
Meter, universal power supply, electrode stand, manual, PC software and USB cable
- Provides automatic documentation for GLP compliance
- AutoRead function improves repeatability
- Easy-to-read backlit display with text menu for easy operation for pH, mV (ORP), ISE and temperature
- Fast data transfer via USB in .csv or ASCII format or the optional integrated printer
- Easy calibration with adjustable calibration timer which indicates when to perform the next calibration routine
- Automatic or manual temperature compensation
- 1, 2, 3, 4 or 5 point calibration for pH
- 1, 2, 3, 4, 5, 6 or 7 point calibration for ISE, also non- linear
- BNC connection for pH, mV and ISEs; banana connection for temperature probes (sold separately)
- Continuous Monitoring Control (CMC) function for monitoring the ideal measuring range for pH measurements based on the last calibration
- Wide selection of electrodes: TruLab analog pH/mV/ISE/temp probes
- Four alkaline AA batteries provide approx.150 hours with illumination; universal power supply; also compatible with four NiMH AA rechargeables
Parameters Measured:
- pH, mV (ORP/Redox)
- Temperature
- ISEs: Ammonia, Ammonium, Bromide, Cadmium, Calcium, Chloride, Copper, Cyanide, Fluoride, Iodide, Lead, Nitrate, Potassium, Silver/Sulfide, Sodium
Specifications:
- Buffer Recognition: Yes
- Sampling: Yes
- User Calibratable: Yes
- Logging Capabilities: Yes
- pH Range: -2.0 to 20.0; -2.00 to 20.00; -2.000 to 19.999
- pH Resolutions: 0.1; 0.01; 0.001
- pH Accuracy (±1 digit): ±0.1; ±0.01; ±0.005 (temp. of test sample 15° to 35°C; within a range of ±2 pH of the calibration point)
- mV Range: -1200 to 1200.0; -2500 to 2500
- mV Resolution: 0.1; 1.0
- mV Accuracy: ±0.3; ±1.0 (temp. of test sample 15° to 35°C)
- ISE Range (mg/L, mg/kg, mmol/L, ppm, %): 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 999999
- ISE Range (μmol/L): 0.000 to 9.999; 10.00 to 99.99; 100.0 to 999.9; 1000 to 9999
- ISE Resolution (mg/L, mg/kg, mmol/L, ppm, %): 0.001; 0.01; 0.1; 1
- ISE Resolution (μmol/L): 0.001; 0.01; 0.1; 1
- ISE Accuracy: ±0.3; ±1.0 (temp of test sample 15° to 35°C)
- Temperature Range: -5° to +105°C (23° to 221°F)
- Temperature Resolution: 0.1
- Temperature Accuracy: ±0.1
- Memory: 5000 data sets auto logging, 500 data sets manual logging
- Manual Temperature Input: -25° to +130°C (-13° to +266°F) in increments of 1
- Storage Temperature: -25° to +65°C (-13° to +149°F)
- Unit of Measure: Sample subtraction, Known addition (double), Sample addition, ISE Measurements: Blank value correction, increment methods: Known addition (single), Known subtraction
- Weight: 0.8kg (1.8 lb.)
- Dimensions (L x W x H): 23 x 19 x 8cm (9 x 7.5 x 3.2 in.)
Order Info
Cables, probes/sensors, and accessories sold separately.
Certifications
IP43, electrical protective class III, cETLus, CE
Compliance
GLP/AQA compliant
Warranty and Services
Three years
Specifications
Specifications
| Accuracy (mV) | ±0.3; ±1.0 (temp. of test sample 15° to 35°C) |
| Accuracy (pH) | ±0.1; ±0.01; ±0.005 (temp. of test sample 15° to 35°C; within a range o |
| Accuracy (Temperature) | ±0.1 |
| Battery Life | approx.150 hours with illumination, 1000 hours without |
| Battery Type | 4 x alkaline AA batteries (also compatible with four NiMH AA rechargeables) |
| Calibration Points (pH) | 1, 2, 3, 4 or 5 point with or without autobuffer recognition; USA and NIST and 22 pre-programmed buf |
| Certifications/Compliance | CE, cETLus; electrical protective class III; IP43 |
| Data Points | 5000 data sets auto logging, 500 data sets manual logging |
| Calibration Points (ISE) | 2 to 7 |
| Range (ISE) | 0.000 to 9.999, 10.00 to 99.99, 100.0 to 999.9, 1000 to 999999 μmol/L, 0.000 to 9.999, 10.00 to 99.99, 100.0 to 999.9, 1000 to 9999 in mg/L, mg/kg, mmol/L, ppm, % |
| Show More |
By clicking Submit, you acknowledge that you may be contacted by Fisher Scientific in regards to the feedback you have provided in this form. We will not share your information for any other purposes. All contact information provided shall also be maintained in accordance with our Privacy Policy.