Promotional price valid on web orders only. Your contract pricing may differ. Interested in signing up for a dedicated account number?
Learn More

Mirus Bio™ Ingenio™ EZporator™ Electroporation System

Catalog No. MIR51000
Click to view available options
:
Ingenio™ EZporator™ Electroporation System; Length: 22 cm; Width: 25 cm

Mirus Bio™ Ingenio™ EZporator™ Electroporation System provides a robust and universal approach for multiple insect and mammalian cell culture applications, including gene expression, RNAi, stable cell line generation, and CRISPR/Cas genome editing.

  • Digital Liquid Crystal Display (LCD) user interface
  • Pulse Generator to deliver exponential decay pulse at user-determined voltage
  • Cuvette chamber compatible with 0.1 cm, 0.2 cm or 0.4 cm cuvettes
  • Power cords with Type B, F and G electrical outlet plugs
  • Technical Specs: Low Voltage (LV) mode and High Voltage (HV) mode
  • Dual monitoring and display of peak output voltage and time constant
  • Streamlined workflow
    • Initialize instrument
    • Set voltage and mode with Voltage Adjust knob
    • Deliver Electroporation Pulse button
  • European regulations and directives: 2014/35/EU Low Voltage Directive (LVD), 2014/30/EU Electromagnetic Compatibility directive (EMC), 2012/19/EU Waste electrical and electronic equipment directive (WEEE), 2011/65/EU Restriction on the use of certain hazardous substances directive (RoHS)
  • Standards to Demonstrate Conformity: EN 61010-1:2010 Safety requirements for electrical equipment for measurement,control and laboratory use, General Requirements EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use
  • EMC requirements: EN 50581:2012 Assessment of products with respect to RoHS
    EMC and Safety testing performed by Intertek Testing Services
  • Listing Number 9801781
  • Test reports: 103276746BOX-001, -002, -003
  • Conforms to UL61010-1:2012
  • Certified to CSA C22.2#61010-1:2012

Specifications

For Use With (Application) Mammalian and insect cell transfection
For Use With (Equipment) Ingenio Electroporation Kits and Solution
Frequency 50/60 Hz
Material Metal Case
Product Type Electroporation System
Voltage 20 to 2500 V
Voltage Range (HV Mode) 30 to 2500 V with 10 V resolution
Voltage Range (LV Mode) 20 to 400 V with 2 V resolution
Display 16-character Backlit LCD
Field Type Exponential decay
Includes Ezporator™ Pulse Generator (quantity 1)​, Ezporator™ Cuvette Chamber (quantity 1)​, Power cords (quantity 3 – Type B, F and G)​
Length (English) 22 cm
Length (Metric) 22 cm
Life Span 1 year warranty
Operating Temperature (English) 50°F to 104°F
Pulse Length Range 5 ms to 125 ms
Operating Temperature (Metric) 10°C to 40°C
Quantity 1 Ea.
Relative Humidity 60%
Width (English) 25 cm
Width (Metric) 25 cm
Electrical Requirements Voltage: 100 to 240 Vac, 50/60 Hz, CAT I; Power: 800 W; Fusion: 2.5 A, T rating 250 V
Sterility Non-sterile
Capacitance (HV Mode) 36 μF
Capacitance (LV Mode) 1050 μF
Connection Type Electrode Connector Cable
Charging Time 5 sec.
Interface Digital LCD
Pulse Form Exponential Decay Wave
Show More Show Less
Product Content Correction

Your input is important to us. Please complete this form to provide feedback related to the content on this product.

Product Title

By clicking Submit, you acknowledge that you may be contacted by Fisher Scientific in regards to the feedback you have provided in this form. We will not share your information for any other purposes. All contact information provided shall also be maintained in accordance with our Privacy Policy.

Your feedback has been submitted: Thank you for helping us improve our website.