Learn More
Thermo Scientific™ SuperSignal™ Western Blot Enhancer
Description
Thermo Scientific™ SuperSignal Western Blot Enhancer contains a membrane treatment reagent and a primary antibody diluent that increase both signal intensity and sensitivity 3- to 10-fold compared to detection performed without it.
Highlights:
Increase sensitivity – achieve 3- to 10-fold increase in signal intensity and sensitivity
Improve specificity – improves signal-to-noise ratio for poor quality and low affinity antibodies
Better clarity – reduces background for cleaner Western blots
Membrane compatibility – provides effective signal enhancement with PVDF and nitrocellulose membranes
Substrate compatibility – validated for use with chromogenic, chemiluminescent and fluorescent detection methods
Specifications
Specifications
| Product Type | Western Blot Enhancer |
| Detection Method | Chemiluminescence, Fluorescence, Colorimetric |
| Form | Liquid |
| For Use With (Application) | Western Blot |
| Membrane Compatibility | Nitrocellulose, PVDF |
| Product Line | SuperSignal |
| Quantity | 50 mL |
| Content And Storage | Upon receipt store solutions at 4°C Products are shipped at ambient temperature. |
Frequently Asked Questions (FAQs)
Here are potential causes for high background and corresponding solutions:
- Too much HRP in the system: Further dilute the HRP-conjugate, see guidelines in Table 1 in the user guide (https://assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0018858_SuperSignalWestAtto_UG.pdf).
- Inadequate blocking: Optimize blocking conditions.
- Inappropriate blocking reagent: Try a different blocking reagent. Use clear milk or StartingBlock blocking buffer instead of dry milk.
- Inadequate washing: Increase length, number or volume of washes.
- Film has been overexposed: Decrease exposure time or use Pierce Background Eliminator (Cat No. 21065).
- Concentration of antigen or antibody is too high: Decrease amount of antigen or antibody.
- Poor antibody specificity: Use SuperSignal Western Blot Enhancer (Cat. No. 46640).
We also recommend minimizing handling of the membrane, using clean gloves while handling, clean plasticware, glassware, forceps, or other contact surfaces. We also recommend trimming of the membrane to remove excess area.
Not necessarily. The SuperSignal Western Blot Enhancer will work with any chemiluminescent substrate (including SuperSignal Western Blot substrates), as well as colorimetric or fluorometric detection systems.
The reagents in both kits are different. Western Blot Signal Enhancer (Cat. No. 21050) is a 2-step pre-treatment that occurs before blocking. It basically relaxes the protein on the membrane making the epitope easier to detect. SuperSignal Western Blot Enhancer (Cat. Nos. 46640, 46641) contains a membrane treatment reagent and a primary antibody diluent that increase both signal intensity and sensitivity 3- to 10-fold compared to a detecion performed without it. We have found that SuperSignal Western Blot Enhancer shows much less background while enhancing signal and sensitivity when compared to Western Blot Signal Enhancer.
The SuperSignal Western Blot Enhancer kit contains two reagents:
- Antigen Pretreatment Solution: used after transferring proteins to the membrane and before blocking or staining the membrane
- The Primary Antibody Diluent: used to dilute the primary antibody
Safety and Handling
For Research Use Only. Not for use in diagnostic procedures.
By clicking Submit, you acknowledge that you may be contacted by Fisher Scientific in regards to the feedback you have provided in this form. We will not share your information for any other purposes. All contact information provided shall also be maintained in accordance with our Privacy Policy.